NEO ECG T180 (ECG TABLET with printer)
Product information
Description
The NEO ECG-T180 is a light and portable ECG tablet designed for the acquisition and display of 9/12/15/18-lead ECG waveforms, as well as heart rate detection. Featuring a 10.1-inch high- resolution color touchscreen, it offers user-friendly operation and versatile functionality.
• 10.1“High-Resolution Color Touchscreen: Easy to operate, portable design.
• Power Supply Options: Can be powered by an external DC power supply, a built-in rechargeable lithium battery, or a charging port.
• Lead Support: Synchronous acquisition and display of 9/12/15/18-lead waveforms, as well as heart rate detection.
Key Features:
• Automatic Detection: Supports automatic pacemaker detection and marking.
• Analysis Functions: Includes automatic analysis, RR analysis, HRV, medication tests, ECG event mode.
• Sampling Modes: Offers pre-sampling, real-time sampling, periodic sampling, and trigger sampling.
• Patient Data Management: Input patient information via virtual alphanumeric keyboard and barcode scanning. Store, preview, review, edit, export, upload, print, and search patient data.
• Data Transmission: Wireless transmission of ECG data via WiFi and mobile networks.
• Printing Options: Built-in thermal printer and support for external laser printers.
• Data Export: Export patient data to USB flash disk via USB connector.
• Security: User login permission control with password or account & password authentication.
• Connectivity: Supports connection with AI-ECG PLATFORM for intelligent diagnosis (optional).
Technical Specifications:
Heart Rate Range: 30bpm~300bpm
Accuracy: ±1
Leads: 9, 12, 15, 18 lead synchronous acquisition
A/D Conversion: 24 bits
Sampling Rate: 32,000 samples/sec
Common Mode Rejection Ratio: ≥140dB (AC filter on), ≥120dB (AC filter off)
Time Constant: ≥5s
Frequency Response: 0.01HZ~350HZ (+0.4db-3.0db)
Sensitivity: Auto, 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV, less than ±5% error
Filters: AC filter: 50Hz, 60Hz, Off; EMG filter: 25Hz, 35Hz, 45Hz, Off; ADS filter: 0.01 Hz, 0.05 Hz, 0.32 Hz, 0.67 Hz; Low pass filter: 75Hz, 100Hz, 150Hz, 300Hz, Off
Paper Speed: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, less than ±3% error
Input Impedance: ≥100MΩ (10Hz)
Input Circuit Current: ≤10nA
Calibration Voltage: 1mV±2%
Depolarization Voltage: ±900mV, ±5%
Noise Level: ≤12.5μV
Amplitude Quantization: 0.95 μV/LSB
Recovery Time
after Defibrillation Discharge: <10s
Pacer Pulse Display: Pacing pulse with amplitude of ±2mV~±700mV, duration of 0.1ms~2.0ms, rise time of less than 100μs, and frequency of 100/min can be displayed on the ECG recording.
Minimum Detectable Signal: 20μVp-p
Analysis Algorithm: Glasgow
Main Unit: 10.1” Tablet, 10.1” Base